UPDF AI

Progress in the Development of Frequency-Domain Multiplexing for the X-ray Integral Field Unit on Board the Athena Mission

H. Akamatsu,L. Gottardi,11 Authors,N. Wakeham

2020 · DOI: 10.1007/s10909-020-02351-3
17 Citations

Abstract

Frequency-domain multiplexing (FDM) is the baseline readout system for the X-ray Integral Field Unit on board the Athena mission. Under the FDM scheme, TESs are coupled to a passive LC filter and biased with alternating current (AC bias) at MHz frequencies. Using high-quality-factor LC filters and room-temperature electronics developed at SRON and low-noise two-stage SQUID amplifiers provided by VTT, we have recently demonstrated good performance with the FDM readout of Mo/Au TES calorimeters with Au/Bi absorbers. We have achieved a performance requested for the demonstration model with the single-pixel AC bias (

ΔE=1.8 eV\Delta E=1.8\,\hbox {eV}

Δ E = 1.8 eV ) and nine-pixel multiplexing (

ΔE=2.6 eV\Delta E=2.6\,\hbox {eV}

Δ E = 2.6 eV ) modes. We have also demonstrated 14-pixel multiplexing with an average energy resolution of 3.3 eV, which is limited by nonfundamental issues related to FDM readout in our laboratory setup.