UPDF AI

Methodology of thermal drift measurements for surface guided radiation therapy systems and clinical impact assessment illustrated on the C-Rad Catalyst+ HD system

J. Lehmann,Therese S. Standen,3 Authors,J. Simpson

2022 · DOI: 10.1016/j.tipsro.2022.02.005
Technical Innovations & Patient Support in Radiation Oncology · 11 Citations

TLDR

Methods to measure the Drift of individual camera pods as well as the drift of the combined clinical signal are described and recipes are provided on how to calculate relevant uncertainties in patient setup and patient position monitoring with SGRT.

Abstract

Thermal drift of optical systems employed for surface guided radiation therapy (SGRT) adds uncertainty to patient setup and monitoring. This work describes methods to measure the drift of individual camera pods as well as the drift of the combined clinical signal. It presents results for four clinical C-Rad Catalyst+ HD systems. Based on the measured clinical drift, recipes are provided on how to calculate relevant uncertainties in patient setup and patient position monitoring with SGRT. Strategies to reduce the impact of drift are explained. While the results are specific to the systems investigated, the methodology is transferable and the clinical recipes are universally applicable.