UPDF AI

Absolute calibration of the spectral responsivity of thermal detectors in the near-infrared (NIR) and mid-infrared (MIR) regions by using blackbody radiation

Tobias T. Pohl,P. Meindl,4 Authors,J. Hollandt

2021 · DOI: 10.5194/JSSS-10-109-2021
6 Citations

Abstract

Abstract. The Physikalisch-Technische Bundesanstalt (PTB) has set up an additional measurement approach for the absolute calibration of the spectral responsivity of detectors in the near-infrared (NIR) and mid-infrared (MIR) spectral range. This alternative method uses the radiation of a blackbody operating at about 1200 K with a precision aperture. The blackbody radiation can be calculated by Planck's law and is additionally spectrally selected by accurately characterized optical bandpass filters. Thus, a calibration of the spectral responsivity of a detector with respect to irradiance can be achieved at the bandpass wavelength of the applied transmission filters. If the aperture of the detector is known, the spectral responsivity can also be calculated with respect to radiant power. Thermopile detectors with known aperture size were calibrated in terms of

their spectral responsivity with several bandpass filters in the spectral

range between 1.5  µ m up to 14  µ m with relative standard

measurement uncertainties between 5 % and 19 %. The obtained results

are consistent with previous calibrations at PTB's national primary detector standard. Therefore, this additional measurement approach is a further validation of the existing primary method which is based on a cryogenic radiometer and extends the usable wavelength range.