UPDF AI

A Feature-based Approach on Contact-less Blood Pressure Estimation from Video Data

Carolin Wuerich,Eva-Maria Humm,Christian Wiede,Gregor Schiele

2022 · DOI: 10.23919/eusipco55093.2022.9909563
European Signal Processing Conference · 10 Citations

Abstract

Conventional blood pressure monitors and sensors have several limitations in terms of accuracy, measurement time, comfort or safety. To address these limitations, we realized and tested a surrogate-based contact-less blood pressure estimation method which relies on a single remote photoplethysmogram (rPPG) captured by camera. From this rPPG signal, we compute 120 features, and perform a sequential forward feature selection to obtain the best subset of features. With a multilayer perceptron model, we obtain a mean absolute error ± standard deviation of MAE 5.50±4.525.50\pm 4.52 mmHg for systolic pressure and 3.73±2.863.73\pm 2.86 mmHg for diastolic pressure. In contrast to previous studies, our model is trained and tested on a data set including normotensive, pre-hypertensive and hypertensive values.